High Frequency ICP® Accelerometer
Model 621B41
High freq. laboratory ceramic shear ICP® accel, 100 mV/g, 0.8 to 20k Hz, 316L SS case, top exit, 10-32 conn.
- Sensitivity: (±5%) 100 mV/g (10.2 mV/(m/s²))
- Frequency Range: (±5%) 144 to 600000 cpm (2.4 to 10000 Hz)
- Frequency Range: (±3dB) 48 to 1200000 cpm (0.8 to 20000 Hz)
- Measurement Range: ±50 g (±490 m/s²)
- Electrical Connector: 10-32 Coaxial Jack
|
PERFORMANCE
|
ENGLISH |
SI |
|
Sensitivity (±5 %)
|
100 mV/g
|
|
|
|
|
Measurement Range
|
±50 g
|
|
|
|
|
Frequency Range (±5 %)
|
144 to 600,000 cpm
|
|
|
|
|
Frequency Range (±10 %)
|
102 to 900,000 cpm
|
|
|
|
|
Frequency Range (±3 dB)
|
48 to 1,200,000 cpm
|
|
|
|
|
Resonant Frequency
|
2,100 kcpm
|
|
|
|
|
Non-Linearity
|
±1 %
|
|
|
|
|
Transverse Sensitivity
|
≤ 5 %
|
|
|
ENVIRONMENTAL
|
|
Overload Limit (Shock)
|
5,000 g pk
|
|
|
|
|
Temperature Range
|
-65 to 250 °F
|
|
|
|
|
Temperature Response
|
See Graph
|
|
|
|
|
Enclosure Rating
|
IP68
|
|
|
ELECTRICAL
|
|
Settling Time (within 1% of bias)
|
≤ 5.0 sec
|
|
|
|
|
Discharge Time Constant
|
≥ 0.2 sec
|
|
|
|
|
Excitation Voltage
|
18 to 28 VDC
|
|
|
|
|
Constant Current Excitation
|
2 to 20 mA
|
|
|
|
|
Output Impedance
|
<150 Ohm
|
|
|
|
|
Output Bias Voltage
|
8 to 12 VDC
|
|
|
|
|
Spectral Noise (10 Hz)
|
30 µg/√Hz
|
|
|
|
|
Spectral Noise (100 Hz)
|
10 µg/√Hz
|
|
|
|
|
Spectral Noise (1 kHz)
|
5 µg/√Hz
|
|
|
|
|
Electrical Isolation (Base)
|
>108 Ohm
|
|
|
PHYSICAL
|
|
Size - Hex
|
11/16 in
|
|
|
|
|
Size - Height
|
1.03 in
|
|
|
|
|
Weight
|
1.06 oz
|
|
|
|
|
Mounting Thread
|
10-32 Female
|
|
|
|
|
Sensing Element
|
Ceramic
|
|
|
|
|
Sensing Geometry
|
Shear
|
|
|
|
|
Housing Material
|
Stainless Steel
|
|
|
|
|
Sealing
|
Welded Hermetic
|
|
|
|
|
Electrical Connector
|
10-32 Coaxial Jack
|
|
|
|
|
Electrical Connection Position
|
Top
|
|
|
OPTIONAL VERSIONS |
|
|
|
Supplied Accessory : Model M081B05 Mounting Stud 10-32 to M6 X 0.75
|
All specifications are at room temperature unless otherwise specified.
|
| NOTES: |
| [1] |
10-32 has no equivalent in S.I. units. |
| [2] |
Conversion Factor 1g = 9.81 m/s². |
| [3] |
The high frequency tolerance is accurate within ±10% of the specified frequency. |
| [4] |
Typical value. |
| [5] |
Typical. |
| [6] |
Zero-based, least-squares, straight line method. |
|
|
|
View Spec Sheet: 621B41_A.pdf
|
|