High Temperature Industrial ICP® Accelerometer
Model HT628F01
Model 628F01 with high temperature range option (+325°F/+163°C)
- Sensitivity: (±5%) 100 mV/g (10.2 mV/(m/s²))
- Frequency Range: (±5%) 144 to 180000 cpm (2.4 to 3000 Hz)
- Frequency Range: (±3dB) 48 to 480000 cpm (0.8 to 8000 Hz)
- Measurement Range: ±50 g (±490 m/s²)
- Electrical Connector: MIL-C-5015
|
PERFORMANCE
|
ENGLISH |
SI |
|
Sensitivity (±5 %)
|
100 mV/g
|
|
|
|
|
Measurement Range
|
±50 g
|
|
|
|
|
Frequency Range (±5 %)
|
144 to 180,000 cpm
|
|
|
|
|
Frequency Range (±10 %)
|
102 to 300,000 cpm
|
|
|
|
|
Frequency Range (±3 dB)
|
48 to 480,000 cpm
|
|
|
|
|
Resonant Frequency
|
1,080 kcpm
|
|
|
|
|
Broadband Resolution (1 to 10 kHz)
|
1 mg
|
|
|
|
|
Non-Linearity
|
±1 %
|
|
|
|
|
Transverse Sensitivity
|
≤ 5 %
|
|
|
ENVIRONMENTAL
|
|
Overload Limit (Shock)
|
1,000 g pk
|
|
|
|
|
Temperature Range
|
-65 to 325 °F
|
|
|
|
|
Temperature Range (Survivable Limit)
|
350 °F
|
|
|
|
|
Temperature Response
|
See Graph
|
|
|
|
|
Enclosure Rating
|
IP68
|
|
|
ELECTRICAL
|
|
Settling Time (within 1% of bias)
|
≤ 3 sec
|
|
|
|
|
Discharge Time Constant
|
≥ 0.2 sec
|
|
|
|
|
Excitation Voltage
|
18 to 28 VDC
|
|
|
|
|
Constant Current Excitation
|
2 to 10 mA
|
|
|
|
|
Output Impedance
|
<500 Ohm
|
|
|
|
|
Output Bias Voltage
|
8 to 12 VDC
|
|
|
|
|
Spectral Noise (10 Hz)
|
50 µg/√Hz
|
|
|
|
|
Spectral Noise (100 Hz)
|
20 µg/√Hz
|
|
|
|
|
Spectral Noise (1 kHz)
|
6 µg/√Hz
|
|
|
|
|
Electrical Protection
|
RFI/ESD
|
|
|
|
|
Electrical Isolation (Case)
|
>108 Ohm
|
|
|
PHYSICAL
|
|
Size - Hex
|
7/8 in
|
|
|
|
|
Size - Height
|
2.06 in
|
|
|
|
|
Weight
|
3.2 oz
|
|
|
|
|
Mounting Thread
|
1/4-28 UNF
|
|
|
|
|
Mounting Torque
|
2 to 5 ft-lb
|
|
|
|
|
Sensing Element
|
Quartz
|
|
|
|
|
Sensing Geometry
|
Shear
|
|
|
|
|
Housing Material
|
Stainless Steel
|
|
|
|
|
Sealing
|
Welded Hermetic
|
|
|
|
|
Electrical Connector
|
MIL-C-5015
|
|
|
|
|
Electrical Connection Position
|
Top
|
|
|
SUPPLIED ACCESSORIES:
|
| Model 081A40 Mounting Stud (1) |
| Model ICS-1 NIST-traceable single-axis amplitude response calibration from 600 cpm (10 Hz) to upper 5% frequency |
|
OPTIONAL VERSIONS |
|
|
|
Supplied Accessory : Model M081A61 Mounting Stud 1/4-28 to M6 X 1 (1) replaces Model 081A40
|
All specifications are at room temperature unless otherwise specified.
|
| NOTES: |
| [1] |
1/4-28 has no equivalent in S.I. units. |
| [2] |
Constant current should be reduced to ≤6 mA when operating sensor above 250°F. |
| [3] |
Conversion Factor 1g = 9.81 m/s². |
| [4] |
Not recommended for use with magnetic mounting base. |
| [5] |
See PCB Declaration of Conformance PS023 for details. |
| [6] |
The high frequency tolerance is accurate within ±10% of the specified frequency. |
| [7] |
Typical. |
| [8] |
Zero-based, least-squares, straight line method. |
|
|
|
View Spec Sheet: HT628F01_C.pdf
|
 |
High-temp PTFE Cable Assembly
Model 053AE010BZ
|
|
 |
Sensor Mounting Pad
Model 080A118
|
|
 |
Indicator / Alarm (for ICP® sensors)
Model 683A101001
|
|
|